Philips Medical Systems (Cleveland) Inc recalls BrightView XCT Model 882482, gamma camera for Single Photon Emission Computed Tomography (SPECT)
- Recall date
- May 25, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2736-2016
- FDA classification
- Class II
- Brand / firm
- Philips Medical Systems (Cleveland) Inc
- Sold / distributed
- Worldwide distribution. US nationwide, Algeria, Argentina, Armenia, Australia, Austria, Belgium, Brazil, Czech Republic, Denmark, Finland France, Gabon, Germany, Greece, India, Indonesia, Ireland, Japan, Korea, Republic of Korea, Republic of Lebanon, Lithuania, Malaysia, Netherlands, New Zealand, N…
Why it was recalled
The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow setup script.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
BrightView XCT Model 882482, gamma camera for Single Photon Emission Computed Tomography (SPECT)
Get recall alerts
Free email alert whenever Philips Medical Systems (Cleveland) Inc has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Philips Medical Systems (Cleveland) Inc