Medical device recalls Moderate risk

Philips Medical Systems (Cleveland) Inc recalls BrightView XCT Model 882482, gamma camera for Single Photon Emission Computed Tomography (SPECT)

Recall date
May 25, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2736-2016
FDA classification
Class II
Brand / firm
Philips Medical Systems (Cleveland) Inc
Sold / distributed
Worldwide distribution. US nationwide, Algeria, Argentina, Armenia, Australia, Austria, Belgium, Brazil, Czech Republic, Denmark, Finland France, Gabon, Germany, Greece, India, Indonesia, Ireland, Japan, Korea, Republic of Korea, Republic of Lebanon, Lithuania, Malaysia, Netherlands, New Zealand, N…

Why it was recalled

The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow setup script.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

BrightView XCT Model 882482, gamma camera for Single Photon Emission Computed Tomography (SPECT)

Get recall alerts

Free email alert whenever Philips Medical Systems (Cleveland) Inc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Philips Medical Systems (Cleveland) Inc