Philips Medical Systems (Cleveland) Inc recalls Brilliance 64, Computed Tomography X-ray system Product Usage: The Brilliance CT 64 scanner is a whole body Computed To…
- Recall date
- April 1, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2116-2016
- FDA classification
- Class II
- Brand / firm
- Philips Medical Systems (Cleveland) Inc
- Sold / distributed
- US Nationwide Distribution in the states of CA, FL, ID, KS, MI, NY,OH,TX
Why it was recalled
Philips Healthcare received reports from the field that certain Brilliance 64 systems running software version 3.5.4 exhibited intermittent swirl-like ring artifacts that may appear on reconstructed images. A patient rescan may be required if the images cannot be used for interpretation due to the swirl-like artifact.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Brilliance 64, Computed Tomography X-ray system Product Usage: The Brilliance CT 64 scanner is a whole body Computed Tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes
Get recall alerts
Free email alert whenever Philips Medical Systems (Cleveland) Inc has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Philips Medical Systems (Cleveland) Inc