Philips Medical Systems (Cleveland) Inc recalls FV-RF (Clinix RF Plus) Product Usage: A Stationary X-ray System, a device intended for radiographic/fluoroscopic examin…
- Recall date
- November 18, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1005-2017
- FDA classification
- Class II
- Brand / firm
- Philips Medical Systems (Cleveland) Inc
- Sold / distributed
- US Nationwide distribution in the following states: AL, CA, FL, GA, HI, IL, IN, KS, LA, MA, MI, MS, NJ, NY, OH, PA, SC, VA, WI, WV, & WY.
Why it was recalled
There's a potential for liquid penetration into the back side of the front control panel of the X-ray diagnostic table, which can cause a short circuit of the terminals of the tilt operation and cause an unintentional tilt of the table.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
FV-RF (Clinix RF Plus) Product Usage: A Stationary X-ray System, a device intended for radiographic/fluoroscopic examinations of various anatomical regions. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.
Get recall alerts
Free email alert whenever Philips Medical Systems (Cleveland) Inc has a new recall — straight from official government data. Unsubscribe anytime.