Medical device recalls Moderate risk

Philips Medical Systems (Cleveland) Inc recalls FV-RF (Clinix RF Plus) Product Usage: A Stationary X-ray System, a device intended for radiographic/fluoroscopic examin…

Recall date
November 18, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1005-2017
FDA classification
Class II
Brand / firm
Philips Medical Systems (Cleveland) Inc
Sold / distributed
US Nationwide distribution in the following states: AL, CA, FL, GA, HI, IL, IN, KS, LA, MA, MI, MS, NJ, NY, OH, PA, SC, VA, WI, WV, & WY.

Why it was recalled

There's a potential for liquid penetration into the back side of the front control panel of the X-ray diagnostic table, which can cause a short circuit of the terminals of the tilt operation and cause an unintentional tilt of the table.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

FV-RF (Clinix RF Plus) Product Usage: A Stationary X-ray System, a device intended for radiographic/fluoroscopic examinations of various anatomical regions. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.