Medical device recalls Moderate risk

Philips Medical Systems, Inc. recalls Pinnacle3 Software Version 10.0, Model 459800200841. Product Usage: The Pinnacle 3 RTP software allows qualified medica…

Recall date
April 23, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1643-2015
FDA classification
Class II
Brand / firm
Philips Medical Systems, Inc.
Sold / distributed
US Nationwide Distribution in the states of MO, FL, OH

Why it was recalled

A problem (the dose computed in planning mode is calculated incorrectly ) has been detected in the Philips Pinnacle3 software version 10.0 that, if it were to re-occur, could pose a risk for patients or users. Specifically, the dose engine is being passed the wrong snout position. The snout position is used to determine the penumbra of the beam which includes the calculation of the source si

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Pinnacle3 Software Version 10.0, Model 459800200841. Product Usage: The Pinnacle 3 RTP software allows qualified medical professionals to enter patient data into the system, use that data to construct a plan for radiation therapy and evaluate the plan. Optionally, the qualified medical personnel may output the plan in an electronic or printed form for use by other systems in the delivery of treatment to a patient.