Philips Medical Systems, Inc. recalls Pinnacle3 Software Version 10.0, Model 459800200841. Product Usage: The Pinnacle 3 RTP software allows qualified medica…
- Recall date
- April 23, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1643-2015
- FDA classification
- Class II
- Brand / firm
- Philips Medical Systems, Inc.
- Sold / distributed
- US Nationwide Distribution in the states of MO, FL, OH
Why it was recalled
A problem (the dose computed in planning mode is calculated incorrectly ) has been detected in the Philips Pinnacle3 software version 10.0 that, if it were to re-occur, could pose a risk for patients or users. Specifically, the dose engine is being passed the wrong snout position. The snout position is used to determine the penumbra of the beam which includes the calculation of the source si
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Pinnacle3 Software Version 10.0, Model 459800200841. Product Usage: The Pinnacle 3 RTP software allows qualified medical professionals to enter patient data into the system, use that data to construct a plan for radiation therapy and evaluate the plan. Optionally, the qualified medical personnel may output the plan in an electronic or printed form for use by other systems in the delivery of treatment to a patient.
Get recall alerts
Free email alert whenever Philips Medical Systems, Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Philips Medical Systems, Inc.