PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 3 M12 System Code: (1) 722063 (2) 722221
- Recall date
- January 13, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1087-2025
- FDA classification
- Class II
- Brand / firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Sold / distributed
- Worldwide distribution - US Nationwide and the countries of Canada, Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Cayman Islands, Chile, China,…
Why it was recalled
Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Azurion 3 M12 System Code: (1) 722063 (2) 722221