Philips Medical Systems recalls DS/US Proton Feature with Pinnacle3 Software Version 10.0.0 and 14.0.0 Radiation Therapy Planning System Model #4598002…
- Recall date
- January 8, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0745-2016
- FDA classification
- Class II
- Brand / firm
- Philips Medical Systems
- Sold / distributed
- US: Nationwide Distribution in the states of FL, OH, and MO.
Why it was recalled
In DS/US proton planning, there is a correction factor used internally to the dose engine that is calculating improperly in some cases. This can result in an incorrect dose calculation.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
DS/US Proton Feature with Pinnacle3 Software Version 10.0.0 and 14.0.0 Radiation Therapy Planning System Model #459800200621
Get recall alerts
Free email alert whenever Philips Medical Systems has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Philips Medical Systems