Medical device recalls Moderate risk

PageWriter TC70 w/o trolley Government Bundle recalled over fire hazard

Recall date
January 9, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Official notice title
Philips North America, LLC recalls PageWriter TC70 w/o trolley Government Bundle, Product 860353, Software revisions up to and including A.07.05.22 to eva…
Recall number
Z-0858-2019
FDA classification
Class II
Brand / firm
Philips North America, LLC
Sold / distributed
Domestic distribution nationwide. Foreign distribution worldwide.

Why it was recalled

Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to melt and/or the device to ignite, which can cause injury to patient or nearby users.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

PageWriter TC70 w/o trolley Government Bundle, Product 860353, Software revisions up to and including A.07.05.22 to evaluate the electrocardiogram of adult and pediatric patients

Get recall alerts

Free email alert whenever Philips North America, LLC has a new recall — straight from official government data. Unsubscribe anytime.