Philips Ultrasound Inc recalls Transducer S8-3t UDI (01)00884838073524 REF 989605431171 Can be used as accessory with: EPIQ 7, Affiniti 70, EPIQ 5, EP…
- Recall date
- August 30, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1843-2020
- FDA classification
- Class II
- Brand / firm
- Philips Ultrasound Inc
- Sold / distributed
- Worldwide distribution - US Nationwide distribution.
Why it was recalled
The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Transducer S8-3t UDI (01)00884838073524 REF 989605431171 Can be used as accessory with: EPIQ 7, Affiniti 70, EPIQ 5, EPIQ CVxi Ultrasound systems - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studies
Get recall alerts
Free email alert whenever Philips Ultrasound Inc has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Philips Ultrasound Inc