Physio-Control Inc recalls LIFEPAK 15 Monitor/Defibrillator service kits Product Usage: The LIFEPAK(R) 15 Monitor/defibrillator (LP15) is a comple…
- Recall date
- February 1, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0863-2019
- FDA classification
- Class I
- Brand / firm
- Physio-Control Inc
- Sold / distributed
- Worldwide Distribution
Why it was recalled
Certain LIFEPAK 15 Monitors/ Defibrillators were reported to experience a lockup condition after a shock was delivered. This condition is defined as a blank monitor display with LED lights on, indicating power on the device, but no response in keypad and device functions.
Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.
What was recalled
LIFEPAK 15 Monitor/Defibrillator service kits Product Usage: The LIFEPAK(R) 15 Monitor/defibrillator (LP15) is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols. The LP15 monitor/defibrillator is intended for use by trained medical personnel in out-of-doors and indoor emergency care settings within the environmental conditions specified in the Operating Instructions. The LP15 monitor/defibrillator is designed to be used during ground transportation except when specified otherwise. Manual mode monitoring and therapy functions are intended for use on adult and pediatric patients. Automated external defibrillation mode is intended for use on patients eight years of age and older.
Get recall alerts
Free email alert whenever Physio-Control Inc has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Physio-Control Inc