Physio-Control, Inc. recalls LIFEPAK(R) defibrillation electrodes, Catalog # 11101-000016, MIN #3202380-006 The defibrillation electrodes are used o…
- Recall date
- October 27, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0186-2018
- FDA classification
- Class I
- Brand / firm
- Physio-Control, Inc.
- Sold / distributed
- Worldwide distribution: US (nationwide) including Puerto Rico and Guam and countries of: Canada, Austria, Belgium, Czech Republic; Denmark, Finland, France, French Polynesia, Germany, Hungary, Ireland, Italy, Latvia, Netherlands, Australia, China, Hong Kong, Japan, Argentina, Mexico, Norway, Panama…
Why it was recalled
The infant artwork on the product instructs incorrect pad placement. Also, the artwork is missing the Type BF Patient connection and High Voltage Warning symbols.
Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.
What was recalled
LIFEPAK(R) defibrillation electrodes, Catalog # 11101-000016, MIN #3202380-006 The defibrillation electrodes are used only with LIFEPAK EXPRESS(R) AED, LIFEPAK CR(R) Plus AED, LIFEPAK 1000 defibrillator, or LIFEPAK 500 Biphasic AED s with a pink connector.
Get recall alerts
Free email alert whenever Physio-Control, Inc. has a new recall — straight from official government data. Unsubscribe anytime.