Henry Schein CardioChek Starter Kit recalled over fire hazard
- Recall date
- December 2, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- Polymer Technology Systems, Inc. recalls Henry Schein CardioChek Starter Kit, Reference Number 4060 HS, 570-1036
- Recall number
- Z-0834-2020
- FDA classification
- Class II
- Brand / firm
- Polymer Technology Systems, Inc.
- Sold / distributed
- The products were distributed US Nationwide. The products were distributed to the following foreign countries: Australia, Austria, Belgium, Bermuda, Canada, China, Chile, Costa Rica, Cayman Islands, Czech Republic, England, Finland, France, Germany, Greece, Hong Kong, India, Iraq, Italy, Jordan, Le…
Why it was recalled
One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Henry Schein CardioChek Starter Kit, Reference Number 4060 HS, 570-1036