Medical device recalls Moderate risk

Polymer Technology Systems, Inc. recalls Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570…

Recall date
January 16, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1372-2019
FDA classification
Class II
Brand / firm
Polymer Technology Systems, Inc.
Sold / distributed
Domestic distribution nationwide.

Why it was recalled

Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being di P832 Chemistry (Diagnostic) CHH 2 Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412