Medical device recalls Moderate risk

PTS Diagnostics CardioChek Plus Professional Analyzer recalled over fire hazard

Recall date
December 2, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Official notice title
Polymer Technology Systems, Inc. recalls PTS Diagnostics CardioChek Plus Professional Analyzer, New analyzer, boxed, with carrying case, Reference Numbers 2700
Recall number
Z-0832-2020
FDA classification
Class II
Brand / firm
Polymer Technology Systems, Inc.
Sold / distributed
The products were distributed US Nationwide. The products were distributed to the following foreign countries: Australia, Austria, Belgium, Bermuda, Canada, China, Chile, Costa Rica, Cayman Islands, Czech Republic, England, Finland, France, Germany, Greece, Hong Kong, India, Iraq, Italy, Jordan, Le…

Why it was recalled

One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

PTS Diagnostics CardioChek Plus Professional Analyzer, New analyzer, boxed, with carrying case, Reference Numbers 2700

Get recall alerts

Free email alert whenever Polymer Technology Systems, Inc. has a new recall — straight from official government data. Unsubscribe anytime.