Medical device recalls Moderate risk

Precision Medical, Inc. recalls Easy Dial Oxygen Regulator, 168708D and 168715D. Eight liter from .0.25 to 8 liters, 15 liter from 0.5 liter to 15 lite…

Recall date
October 8, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0272-2016
FDA classification
Class II
Brand / firm
Precision Medical, Inc.
Sold / distributed
Distributed in the US to the states of PA, MA, MO, MT, ID, MI, MN, ND, KY, NY, NC, CA, AL, AZ, LA, OH, TX, GA, IN, and OK.

Why it was recalled

These lots of Oxygen Dial Regulators could have the settings label off by one position. When the indicator arrow points to OFF position, the regulator is actually on the half liter setting (0.50), for the 15 liter. For the 8 liter, instead of being in the OFF position, the regulator will be in the quarter liter setting (0.25).

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Easy Dial Oxygen Regulator, 168708D and 168715D. Eight liter from .0.25 to 8 liters, 15 liter from 0.5 liter to 15 liters of oxygen per minute.

Get recall alerts

Free email alert whenever Precision Medical, Inc. has a new recall — straight from official government data. Unsubscribe anytime.