Medical device recalls

Primus Corporation dba Trinity Biotech recalls A. MENARINI diagnostics Hb 9210 PREMIER, LEVEL1, LEVEL II, Hemoglobin A1c Controls for hemoglobin A1c Assay, REF 45052,…

Recall date
October 5, 2021
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0349-2023
FDA classification
Class III
Brand / firm
Primus Corporation dba Trinity Biotech
Sold / distributed
Turkey, India, Romania, Thailand, Vietnam, S. Korea, Peru, Pakistan Austria, UAE, Chile, Italy.

Why it was recalled

The Control Level I material value was showing results exceeding the upper limit of the acceptable range after a period of three (3) months. The unacceptable Control Level 1 value results in the Analyzer generating an error code which stops further testing and nullifies associated patient samples, resulting in a delay in patient results and a potential delay in therapeutic modifications.

Class III means the product violates regulations but is not likely to cause harm.

What was recalled

A. MENARINI diagnostics Hb 9210 PREMIER, LEVEL1, LEVEL II, Hemoglobin A1c Controls for hemoglobin A1c Assay, REF 45052, part number 05-01-0926, A1c analysis

Get recall alerts

Free email alert whenever Primus Corporation dba Trinity Biotech has a new recall — straight from official government data. Unsubscribe anytime.