Medical device recalls Moderate risk

Princeton Biomeditech Corp recalls Osom Ultra hCG Combo Test- IVD immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in serum…

Recall date
August 23, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0259-2025
FDA classification
Class II
Brand / firm
Princeton Biomeditech Corp
Sold / distributed
US Distribution to MO only.

Why it was recalled

This recall was initiated due to customer complaints of false negative of hCG test results when running patient urine samples. Use of the above listed product lot may result in false negative results. An end user reported that the OSOM Ultra hCG Combo Test, lot 563L13, produced false negative test results from 2 patient urine samples as confirmed by quantitative hCG and ultrasound. During further root cause testing and investigation, an additional 6 devices were found to produce false negative test results from hCG positive samples.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Osom Ultra hCG Combo Test- IVD immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in serum or urine for the early confirmation of pregnancy. Model/Catalog Number: 1004