Medical device recalls Moderate risk

Prytime Medical Devices, Inc. recalls PRYTIME MEDICAL REBOA Catheter Introducer Kit, REF: KT1835C, Contents: Needle, 18 GA x 7 CM ECHOGENIC, STERILE, 1 EA, S…

Recall date
December 14, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0959-2021
FDA classification
Class II
Brand / firm
Prytime Medical Devices, Inc.
Sold / distributed
US Nationwide distribution including in the states of NJ, NY, CA, TX, MI, MO, CO, MI, VA, DE, TN, MO, MT, AZ, FL, WI, SC, WA, TX, CT, IN, PA, ME, WI, MA, GA, OH, AR, UT, NY OUS: None

Why it was recalled

There is a potential that a catheter convenience set contains a 21G access needle instead of the correct 18G needle. If the 21G needle is used to gain vascular access then the guidewire included in the kit will not pass through the access needle to facilitate insertion of an introducer sheath and could therefore cause procedure delay or injury.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

PRYTIME MEDICAL REBOA Catheter Introducer Kit, REF: KT1835C, Contents: Needle, 18 GA x 7 CM ECHOGENIC, STERILE, 1 EA, SYRINGE, 30CC, STERILE 1 EA, INTRODUCER SHEATH KIT, 7 FR W/.035 SHORT GUIDEWIRE, STERILE 1 EA., PREFILLED, STERILE FIELD SALINE SYRINGE, 10 ML, 4 EA, CLAMP, 5 FR, STERILE, 1 EA, SCALPEL, #11SS, STERILE, 1 EA, SUTURE, NYLON, 18", 2-0, REVERSE CUTTING, STERILE 1 EA, THREE QUARTER DRAPE, STERILE 1 EA, INSERT, SALINE 1 EA, INSERT, CLAMP 1 EA UDI: (01) 0 0863092 00010 1