Medical device recalls Moderate risk

QiG Group LLC recalls Algostim LLC, Algovita Implanted Pulse Generator stimulator, Model 2412 ( 2 x 12 channel)

Recall date
January 26, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1193-2016
FDA classification
Class II
Brand / firm
QiG Group LLC
Sold / distributed
Distributed in Germany and Luxembourg.

Why it was recalled

The implantable pulse generator (IPG) model 2408 had suboptimal performance establishing a connection to the external devices unless they were brought into close proximity (as close as 5 cm). The following functions may not be available: Reprogramming the stimulator, Adjusting stimulation parameters (on/off, intensity modifications, stimulation program selection), and Recharging the stimulator.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Algostim LLC, Algovita Implanted Pulse Generator stimulator, Model 2412 ( 2 x 12 channel)

Get recall alerts

Free email alert whenever QiG Group LLC has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: QiG Group LLC