Medical device recalls High risk

QUIDEL CARDIOVASCULAR INC recalls Quidel Triage Cardiac Panel: REF: 97000HS, 97021HS (Troponin I), 97022HS (CK-MB and Troponin I), 97000HSEU, 97000HSEUJP…

Recall date
May 25, 2023
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1974-2023
FDA classification
Class I
Brand / firm
QUIDEL CARDIOVASCULAR INC
Sold / distributed
US: TX, NC, CA, LA, SC, OR, HI, OK, AZ, AL, KS, WA, FL, VA, IA, WV, MT, UT, CO MS, NM, MA, NJ, MO AK, IL, MN, WI, IN, GA, NE, MD. OUS: GBR, ITA, FRA, DEU, MCO, EGY, LCA, CHE, ASM, VIR, IRL

Why it was recalled

Raw material issue may cause cardiac panel products to have an average of 30% troponin signal reduction across measurement range 0.05 ng/mL to 30 ng/mL, QC testing won't detect anomaly, so an alternate method is recommended, if unavailable, use recommendations to mitigate potential patient impact, which could be missed/delayed myocardial infarction diagnosis.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

Quidel Triage Cardiac Panel: REF: 97000HS, 97021HS (Troponin I), 97022HS (CK-MB and Troponin I), 97000HSEU, 97000HSEUJP, 97000HZ01, 97000QIL. Lots T13664 - T14045.

Get recall alerts

Free email alert whenever QUIDEL CARDIOVASCULAR INC has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: QUIDEL CARDIOVASCULAR INC