Medical device recalls Moderate risk

R & D Systems, Inc. recalls Quantikine IVD Human sTfR Immunoassay (Human sTransferrin ELISA Kit (sTfR immunological test system)). An enzyme linked…

Recall date
January 20, 2021
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1255-2021
FDA classification
Class II
Brand / firm
R & D Systems, Inc.
Sold / distributed
USA: (FL, LA, CA, OH), CHINA, UNITED KINGDOM, JAPAN, SINGAPORE, Jakarta Selatan,

Why it was recalled

R&D Systems, Inc. received two complaints that the standard signal was low and controls were not within the specification range provided with the kit. Investigation confirmed low standard signal and controls running out of specification for kits returned from the customer and the retained kits.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Quantikine IVD Human sTfR Immunoassay (Human sTransferrin ELISA Kit (sTfR immunological test system)). An enzyme linked immunosorbent assay (ELISA) for the quantitative determination of Soluble Transferrin Receptor (sTfR) concentration in human serum and plasma as an aid in the diagnosis of anemia and polycythemia.

Get recall alerts

Free email alert whenever R & D Systems, Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: R & D Systems, Inc.