R & D Systems, Inc. recalls R&D Systems, Quantikine¿ IVD¿sTfR Immunoassay Kit, for the quantitative determination of soluble transferrin receptor (…
- Recall date
- August 24, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2887-2016
- FDA classification
- Class II
- Brand / firm
- R & D Systems, Inc.
- Sold / distributed
- US: CA, CO, KY, LA, MD, MA, NJ, OH, RI, UT, VA, WA. OUS: AUSTRALIA, China, INDIA, INDONESIA, ISRAEL, JAPAN, MEXICO, SOUTH KOREA, UNITED KINGDOM.
Why it was recalled
R&D Systems, Inc. received two customer complaints that results of the Controls provided with the Quantikine¿ IVD¿ Human sTfR Immunoassay kit were out of the range high. Internal testing within R&D Systems confirmed a problem with Control results being out of range high.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
R&D Systems, Quantikine¿ IVD¿sTfR Immunoassay Kit, for the quantitative determination of soluble transferrin receptor (sTfR) concentrations in human serum and plasma For in Vitro Diagnostic Use, REF DTFR1.
Get recall alerts
Free email alert whenever R & D Systems, Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: R & D Systems, Inc.