Ra Medical Systems, Inc. recalls DABRA Laser (RA-308 Excimer Laser) Intended Use: For use in ablating a channel in occlusive peripheral vascular disease…
- Recall date
- October 16, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0569-2021
- FDA classification
- Class II
- Brand / firm
- Ra Medical Systems, Inc.
- Sold / distributed
- US Distribution to states of: AL, AZ, CA, FL, IL, IN, KY, LA, MD, NC, NV, NY, OK, OR, TX, and WA; and to OUS country: Japan.
Why it was recalled
Due to footswitch not meeting required specification for protection from ingress of solids or liquids. If spilling of liquids near or on the footswitch, it may fail in the 'on' position and allow the laser to continue to lase.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
DABRA Laser (RA-308 Excimer Laser) Intended Use: For use in ablating a channel in occlusive peripheral vascular disease. Indications: For crossing chronic total occlusions in patients with symptomatic infrainguinal lower extremity vascular disease.
Get recall alerts
Free email alert whenever Ra Medical Systems, Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Ra Medical Systems, Inc.