Medical device recalls Moderate risk

Radiometer America Inc recalls AutoCheck5+, Level 2, specifically designed for Radiometer Americas blood gas analyzers The QUALICHECK+ quality contro…

Recall date
September 7, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-3237-2017
FDA classification
Class II
Brand / firm
Radiometer America Inc
Sold / distributed
Worldwide Distribution - US (nationwide) Internationally to Canada, Netherlands, China, Czech Republic, Denmark, Germany, Spain France, Hungary, India, Denmark, Tokyo, South Korea, United Kingdom, Australia, Poland, South Africa, Singapore, Switzerland, Turkey, Norway, Austria, Italy, Sweden, and F…

Why it was recalled

Radiometer has recently become aware that a few of the AUTOCHECK 5+ level 2 ampoules, will give incorrect results for pCO2 and pO2. A defective QC ampoule will for: " pCO2 give a result close to or just outside of the low end of the control range. (4.92  5.85 kPa or 36.9  46.9 mmHg) " pO2 give a considerable higher result than the expected assigned value of 13.9 -14.3 kPa (104-107 mmHg)

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

AutoCheck5+, Level 2, specifically designed for Radiometer Americas blood gas analyzers The QUALICHECK+ quality control (QC) ampoules are specifically designed for Radiometer America s blood gas analyzers