Medical device recalls Moderate risk

Radiometer Medical ApS recalls ABL90 FLEX PLUS Analyzer, Model Number 393-092 - Product Usage: It is intended for use in a laboratory environment, nea…

Recall date
February 3, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1761-2020
FDA classification
Class II
Brand / firm
Radiometer Medical ApS
Sold / distributed
Worldwide distribution - US Nationwide distribution including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of Canada,…

Why it was recalled

The firm received reports of occurrences where the barcode reader misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in connection with a sample measurement. This could result is patient mixup and/or lost of sample, resulting in delayed medical treatment.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

ABL90 FLEX PLUS Analyzer, Model Number 393-092 - Product Usage: It is intended for use in a laboratory environment, near patient or point-of-care setting.