Randox Laboratories Ltd. recalls Chemistry QC Premium Plus, Cat. No. LAL4213, GTIN 05055273209006 - Product Usage: This product is intended for in vitro…
- Recall date
- February 18, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1547-2020
- FDA classification
- Class II
- Brand / firm
- Randox Laboratories Ltd.
- Sold / distributed
- Worldwide distribution - US Nationwide distribution in the states of WV, IN, VA, ID, PR, DE, and countries of West Indies, Czech Republic, France, Germany, Hong Kong, Ireland, Italy, Norway, Philippines, South Africa, South Korea, Sweden, Taiwan (R.O.C.), Thailand, UAE, UK.
Why it was recalled
The control target and range value for Sodium using the ISE indirect method has been assigned incorrectly.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Chemistry QC Premium Plus, Cat. No. LAL4213, GTIN 05055273209006 - Product Usage: This product is intended for in vitro diagnostic use, in the quality control of diagnostic assays. The device is for the control of accuracy.