Medical device recalls

Randox Laboratories Ltd. recalls Evidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drug molecule and…

Recall date
April 5, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1786-2024
FDA classification
Class III
Brand / firm
Randox Laboratories Ltd.
Sold / distributed
US Nationwide distribution.

Why it was recalled

Foam gasket on the Evidence MultiSTAT chip heater assembly may not perform as expected. Foam that is not functioning as expected may cause light to leak into the Charged Coupled Device (CCD) Camera during imaging. This will affect the testing and may delay in reporting results.

Class III means the product violates regulations but is not likely to cause harm.

What was recalled

Evidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drug molecule and metabolites of drugs in human urine Catalog Number: EV4115

Get recall alerts

Free email alert whenever Randox Laboratories Ltd. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Randox Laboratories Ltd.