Medical device recalls Moderate risk

Randox Laboratories Ltd. recalls Lactate Dehydrognease P-L, (UV), Catalog Number LD3818 - Product Usage: device intended for the quantitative in vitro d…

Recall date
August 11, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2963-2020
FDA classification
Class II
Brand / firm
Randox Laboratories Ltd.
Sold / distributed
US Nationwide distribution including in the states of WV, and Puerto Rico.

Why it was recalled

An update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Lactate Dehydrogenase assay on RX Instruments. Additional pipette washes can be implemented as described in the technical bulletin. Interference to the Lactate Dehydrogenase (LOH) reagent would be observed as inconsistencies in Quality Control recovery, which may lead to a delay in running patient samples, or erroneous elevated test results.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Lactate Dehydrognease P-L, (UV), Catalog Number LD3818 - Product Usage: device intended for the quantitative in vitro determination of Lactate Dehydrogenase (LD) activity in serum and plasma. This product is suitable for use on the RX series instruments which includes the RX daytona and RX imola.