Randox Laboratories Ltd. recalls Lactate Dehydrognease P-L, (UV), Catalog Number LD3818 - Product Usage: device intended for the quantitative in vitro d…
- Recall date
- August 11, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2963-2020
- FDA classification
- Class II
- Brand / firm
- Randox Laboratories Ltd.
- Sold / distributed
- US Nationwide distribution including in the states of WV, and Puerto Rico.
Why it was recalled
An update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Lactate Dehydrogenase assay on RX Instruments. Additional pipette washes can be implemented as described in the technical bulletin. Interference to the Lactate Dehydrogenase (LOH) reagent would be observed as inconsistencies in Quality Control recovery, which may lead to a delay in running patient samples, or erroneous elevated test results.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Lactate Dehydrognease P-L, (UV), Catalog Number LD3818 - Product Usage: device intended for the quantitative in vitro determination of Lactate Dehydrogenase (LD) activity in serum and plasma. This product is suitable for use on the RX series instruments which includes the RX daytona and RX imola.
Get recall alerts
Free email alert whenever Randox Laboratories Ltd. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Randox Laboratories Ltd.