Randox Laboratories Ltd. recalls Randox CALCIUM (Ca) Colorimetric Method RX Series Cat. No. CA 3871, R1. Arsenazo Reagent 9 x 51ml GTIN: 05055273200904
- Recall date
- December 4, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0839-2019
- FDA classification
- Class II
- Brand / firm
- Randox Laboratories Ltd.
- Sold / distributed
- WV, PR
Why it was recalled
Update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Calcium assay on RX instruments. Carryover to the Calcium reagent results may result in falsely elevated or decreased results to both patient and QC samples.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Randox CALCIUM (Ca) Colorimetric Method RX Series Cat. No. CA 3871, R1. Arsenazo Reagent 9 x 51ml GTIN: 05055273200904
Get recall alerts
Free email alert whenever Randox Laboratories Ltd. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Randox Laboratories Ltd.