Medical device recalls Moderate risk

Randox Laboratories Ltd. recalls Randox CALCIUM (Ca) COLORIMETRIC METHOD RX DAYTONA PLUS Cat. No. CA 8309 R1, Arsenazo Reagent 4 x 20 ml GTIN: 050552732…

Recall date
December 4, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0840-2019
FDA classification
Class II
Brand / firm
Randox Laboratories Ltd.
Sold / distributed
WV, PR

Why it was recalled

Update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Calcium assay on RX instruments. Carryover to the Calcium reagent results may result in falsely elevated or decreased results to both patient and QC samples.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Randox CALCIUM (Ca) COLORIMETRIC METHOD RX DAYTONA PLUS Cat. No. CA 8309 R1, Arsenazo Reagent 4 x 20 ml GTIN: 05055273208368