Medical device recalls Moderate risk

Randox Laboratories Ltd. recalls RX Imola-IVD Clinical Chemistry Analyzer Catalog Number: RX4900. RX4900R (refurbished)

Recall date
September 6, 2023
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0200-2024
FDA classification
Class II
Brand / firm
Randox Laboratories Ltd.
Sold / distributed
US Nationwide distribution including Puerto Rico

Why it was recalled

RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

RX Imola-IVD Clinical Chemistry Analyzer Catalog Number: RX4900. RX4900R (refurbished)