Randox Laboratories Ltd. recalls The Randox Serology I Positive Control contains the following analytes in human based plasma: Anti HIV 1/2, Anti HTLV 1…
- Recall date
- August 1, 2022
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1772-2022
- FDA classification
- Class II
- Brand / firm
- Randox Laboratories Ltd.
- Sold / distributed
- US Nationwide distribution in the states of WV, NC.
Why it was recalled
***Update** :HBsAg analyte is also Non-Reactive on the Ortho Vitros analyser as well as the Beckman Coulter DxI. Randox Serology I Positive Control will test negative (Non-Reactive) for Marker HBsAg resulting in delay in reporting patient results due to the positive control O36SR testing negative (Non-Reactive) for Marker HBsAg
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
The Randox Serology I Positive Control contains the following analytes in human based plasma: Anti HIV 1/2, Anti HTLV 1/2, Anti HCV, HBsAg, Anti HBc, and Anti Treponema pallidum IgG. The product is delivered as a liquid in a kit of 3 identical ready-to-use vials of 5 mL each.