RAYSEARCH LABORATORIES AB recalls Radiation Therapy Treatment Planning System, Model 5.0
- Recall date
- May 30, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0079-2017
- FDA classification
- Class II
- Brand / firm
- RAYSEARCH LABORATORIES AB
- Sold / distributed
- AZ, CA, FL, MT, NC NY, TX, OH & WA
Why it was recalled
For a treatment plan consisting of multiple beam sets, the table for ROI plan dose statistics in the report may show the statistics for a beam set dose. The error can only be triggered when using a report template where statistics for a beam set dose is included as the last dose statistics prior to the plan dose statistics.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Radiation Therapy Treatment Planning System, Model 5.0