Medical device recalls Moderate risk

RAYSEARCH LABORATORIES AB recalls RaySearch RayStation 2.5, 3.0, 3.5 and 4.0; Radiation Therapy Treatment Product Usage: RayStation is a software system…

Recall date
May 8, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1679-2015
FDA classification
Class II
Brand / firm
RAYSEARCH LABORATORIES AB
Sold / distributed
US Nationwide Distribution

Why it was recalled

RaySearch Laboratories AB became aware of an issue with the dose calculation for some imported VMAT plans where controls points are defined with wide gantry angle spacing.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

RaySearch RayStation 2.5, 3.0, 3.5 and 4.0; Radiation Therapy Treatment Product Usage: RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs. The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.

Get recall alerts

Free email alert whenever RAYSEARCH LABORATORIES AB has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: RAYSEARCH LABORATORIES AB