RAYSEARCH LABORATORIES AB recalls RayStation 4.7, version 4.7.1, used with Structure Definition or Automatic Breast Planning. Radiation Therapy Treatment…
- Recall date
- May 27, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2457-2015
- FDA classification
- Class II
- Brand / firm
- RAYSEARCH LABORATORIES AB
- Sold / distributed
- Distributed in CO, IL, MI, NC, OH,TN, and WA.
Why it was recalled
Issue with the algorithm for ROI contraction in RayStation 4.7 when non-uniform contraction distances are used. The ROI contraction tool uses six distances as input: right/left, inferior/superior and posterior/anterior. These contraction distances are, in error, pairwise reversed, i.e. right is interpreted as left, inferior is interpreted as superior, and posterior is interpreted as anterior.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
RayStation 4.7, version 4.7.1, used with Structure Definition or Automatic Breast Planning. Radiation Therapy Treatment Planning System, for treatment planning and analysis of radiation therapy.
Get recall alerts
Free email alert whenever RAYSEARCH LABORATORIES AB has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: RAYSEARCH LABORATORIES AB