Medical device recalls High risk

Respironics California Inc recalls Philips V60 Ventilators, Material No. 850008, 1053613, 1053614, 1053615, 1053616, 1053617, 1076715, 1076716, 1076717, D…

Recall date
April 24, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2061-2017
FDA classification
Class I
Brand / firm
Respironics California Inc
Sold / distributed
Worldwide Distribution- US (nationwide) and countries of: Argentina, Australia, Austria, Bahrain, Barbados, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Hong Kong, Iceland, India, Indonesia,…

Why it was recalled

Over time, low-frequency vibrations can cause the pins within the female connectors on the Motor Controller to Data Acquisition Board ribbon cable to become partially displaced causing momentary high resistance that prevents or restricts data travelling through the cable during the high resistance event.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

Philips V60 Ventilators, Material No. 850008, 1053613, 1053614, 1053615, 1053616, 1053617, 1076715, 1076716, 1076717, DU1053617, U1053617. Old Material No. 850008, 989805628251, 989805612101, 989805613391, 989805613661, 989805611761, 989805627411, 989805627431, 989805627441, 989805616411, 989805636631.

Get recall alerts

Free email alert whenever Respironics California Inc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Respironics California Inc