Medical device recalls Moderate risk

Riverpoint Medical, LLC recalls Iconix Knotless 1.4mm Anchor with XBraid S (Blue/Black) REF 3911-714-520 Iconix Knotless 1.4mm Anchor with XBraid S (Bl…

Recall date
October 25, 2023
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0518-2024
FDA classification
Class II
Brand / firm
Riverpoint Medical, LLC
Sold / distributed
U.S. Nationwide distribution in the state of CA.

Why it was recalled

Use of a 1.4 drill in hard bone conditions caused increased insertion forces leading to subsequent failures of the inserter, when the inserters bend, it may also move off the drill hole location leading to damage of the suture construct during insertion. Suture damage may lead to suture breaks during use.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Iconix Knotless 1.4mm Anchor with XBraid S (Blue/Black) REF 3911-714-520 Iconix Knotless 1.4mm Anchor with XBraid S (Black/White) REF 3911-714-521 The Iconix Knotless anchors are intended to be used for soft-tissue to bone fixation in the foot, ankle, knee, hip, hand, wrist, elbow and shoulder.

Get recall alerts

Free email alert whenever Riverpoint Medical, LLC has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Riverpoint Medical, LLC