Roche Diabetes Care, Inc. recalls Accu-Chek Performa Test Strip, Blood Glucose Monitoring System, Model number 07299702001
- Recall date
- July 28, 2021
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2352-2021
- FDA classification
- Class II
- Brand / firm
- Roche Diabetes Care, Inc.
- Sold / distributed
- Worldwide distribution - US Nationwide distribution.
Why it was recalled
Test strip vials may open while inside sealed cartons during shipment. An open vial might expose the test strips to humidity, which might damage the strips and result in inaccurate results, potentially leading to inappropriate therapy decisions and adverse health consequences.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Accu-Chek Performa Test Strip, Blood Glucose Monitoring System, Model number 07299702001