Medical device recalls Moderate risk

Roche Diagnostics Corporation recalls Sample probes associated with: cobas 6000 Series System, cobas 8000 Modular Analyzer Series, cobas c 311 analyzer cobas…

Recall date
March 21, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2049-2017
FDA classification
Class II
Brand / firm
Roche Diagnostics Corporation
Sold / distributed
US Nationwide.Distribution

Why it was recalled

Due to a design change in 2015, corrosion on the sample probe electrical contacts can occur and lead to a disturbance of the sample liquid level detection (LLD). When disturbance of the sample liquid level detection (LLD) occurs, the affected sample probe may dip into the sample material deeper than intended. This may lead to carryover and medical risk cannot be excluded.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Sample probes associated with: cobas 6000 Series System, cobas 8000 Modular Analyzer Series, cobas c 311 analyzer cobas c 501/502 cobas c 701/702 cobas 8000 ISE Product Usage: The cobas 6000 series is a fully automated, random-access, software-controlled system for immunoassay and photometric analysis intended for qualitative and quantitative in vitro determinations using a wide variety of tests. It is optimized for high throughput workloads in the professional environment using a combination of Ion selective electrodes (ISE), a photometric analysis unit an immunoassay analysis module. The cobas 8000 Modular Analyzer Series is a fully automated system for Clinical chemistry analysis, intended for the in vitro qualitative and quantitative determination of analytes in body fluids. It is optimized for high throughput workloads in the professional environment using a combination of ion selective electrodes (ISE) and a photometric analysis unit. The cobas c 311 analyzer is an in-vitro diagnostic analyzer capable of performing clinical chemistry, specific protein and electrolyte tests. Analytes are measured photometrically or turbidimetrically; the analyzer also has an optional ISE module for measuring sodium, potassium and chloride.