Roche Diagnostics Corporation recalls The Tina-quant lgA Gen.2 is an immunoturbidimetric assay. Anti-lgA antibodies react with antigen in the sample to form…
- Recall date
- June 3, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1979-2019
- FDA classification
- Class II
- Brand / firm
- Roche Diagnostics Corporation
- Sold / distributed
- The products were distributed to the following US states: AL, IN, and KY.
Why it was recalled
Calibration failures with Std.E, Sens.E, or Dup.E error due to atypical reaction kinetics and quality control failures have occurred due to contamination of R1 reagent with R3 reagent.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
The Tina-quant lgA Gen.2 is an immunoturbidimetric assay. Anti-lgA antibodies react with antigen in the sample to form an antigen/antibody complex which is measured turbidimetrically.
Get recall alerts
Free email alert whenever Roche Diagnostics Corporation has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Roche Diagnostics Corporation