Cobas p 471 centrifuge unit calculator/data processing module recalled over laceration hazard
- Recall date
- July 15, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- Roche Diagnostics Operations, Inc. recalls cobas p 471 centrifuge unit calculator/data processing module, for clinical use Product Usage: The cobas p312 pre-analy…
- Recall number
- Z-0108-2017
- FDA classification
- Class II
- Brand / firm
- Roche Diagnostics Operations, Inc.
- Sold / distributed
- US Nationwide Distribution
Why it was recalled
Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
cobas p 471 centrifuge unit calculator/data processing module, for clinical use Product Usage: The cobas p312 pre-analytical system is a computer controlled stand-alone system for sorting and decapping sample tubes. As an option, the system works with an online connection.