Medical device recalls Moderate risk

Roche Diagnostics Operations, Inc. recalls LDL-Cholesterol gen.2. The Boehringer Mannheim Direct LDL-Cholesterol test in intended for the quantitative determinati…

Recall date
May 7, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1927-2015
FDA classification
Class II
Brand / firm
Roche Diagnostics Operations, Inc.
Sold / distributed
US (nationwide) including PR.

Why it was recalled

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

LDL-Cholesterol gen.2. The Boehringer Mannheim Direct LDL-Cholesterol test in intended for the quantitative determination of low-density lipoprotein Cholesterol (LDL-C) in serum and plasma.