Roche Diagnostics Operations, Inc. recalls Triglycerides GPO. In vitro test for the quantitative determination of triglycerides in human serum and plasma on the C…
- Recall date
- May 7, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1929-2015
- FDA classification
- Class II
- Brand / firm
- Roche Diagnostics Operations, Inc.
- Sold / distributed
- US (nationwide) including PR.
Why it was recalled
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Triglycerides GPO. In vitro test for the quantitative determination of triglycerides in human serum and plasma on the COBAS Integra 800 and 400 plus systems, cobas c111, Roche/Hitachi MODULAR systems, and cobas c systems.
Get recall alerts
Free email alert whenever Roche Diagnostics Operations, Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Roche Diagnostics Operations, Inc.