Roche Molecular Systems, Inc. recalls cobas EGFR Mutation Test V2 Materials number CE-IVD: EGFR v2: 07248563190 cfDNA: 07247737190 Device Identifier: EGFR v2…
- Recall date
- April 5, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2254-2016
- FDA classification
- Class II
- Brand / firm
- Roche Molecular Systems, Inc.
- Sold / distributed
- Foriegn Distribution Austria, Belgium, Croatia, Czech Republic, Denmark, France, Germany, Greece, Hungary , Italy, Lithuania, Luxembourg, Netherlands, Norway, Poland, Romania, Slovakia, Slovenia, Spain, Switzerland, Turkey, United Kingdom cfDNA: Austria, Belgium, Croatia, Czech Republic, Denmark, F…
Why it was recalled
During internal studies using contrived plasma specimens, several mutations (L8568R, Exon 19 deletion, T790M) inconsistently generated "No Mutation Detected" (ie False Negative ) result with the cobas EGFR mutation Text V2.0 when utilizing the cobas cfDNA Sample Preparation kit.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
cobas EGFR Mutation Test V2 Materials number CE-IVD: EGFR v2: 07248563190 cfDNA: 07247737190 Device Identifier: EGFR v2.0: 00875197005448 cfDNA: 00875197005424 Product Usage: A real time PCR Test for the qualitative detection and identifications of mutations n exons 18, 19, 20 and 21 of the epidermal growth factor receptor (EGFR) gene in DNA derived from formalin-fixed paraffin-embedded (FFPET) tumor tissue or plasma from non-small cell lung cancer (NSCLC) patients. The test also is intended to be used as an aid in selecting patients with NSCLC for therapy with an EGFR tyrosine kinase inhibitor (TKI).
Get recall alerts
Free email alert whenever Roche Molecular Systems, Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Roche Molecular Systems, Inc.