Medical device recalls Moderate risk

Roche Molecular Systems, Inc. recalls cobas Influenza A/B & RSV UC (Utility Channel) Qualitative nucleic acid test for use on the cobas 6800/8800 Systems; Ma…

Recall date
December 21, 2022
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1106-2023
FDA classification
Class II
Brand / firm
Roche Molecular Systems, Inc.
Sold / distributed
Domestic distribution nationwide. Foreign distribution worldwide.

Why it was recalled

The firm received customer complaints regarding false negative Influenza A (Flu A) results and late Flu A Target Ct values. A false negative influenza A result may lead to additional testing, psychological distress, and delays to targeted therapy for influenza as well as true diagnosis.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

cobas Influenza A/B & RSV UC (Utility Channel) Qualitative nucleic acid test for use on the cobas 6800/8800 Systems; Material number 09233962190