Medical device recalls Moderate risk

Roche Molecular Systems, Inc. recalls MagNA Pure 96 Instrument, clinical sample concentrator, Material number 06541089001 - Product Usage: This protocol is i…

Recall date
June 3, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2551-2020
FDA classification
Class II
Brand / firm
Roche Molecular Systems, Inc.
Sold / distributed
Worldwide distribution - US Nationwide distribution.

Why it was recalled

When using Sample Transfer protocol version 3.0, the drop catcher is not activated on the MagNA Pure 96 instrument, which introduces a potential risk for cross-contamination leading to erroneous results.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

MagNA Pure 96 Instrument, clinical sample concentrator, Material number 06541089001 - Product Usage: This protocol is intended for pipetting samples into a MagNA Pure 96 Processing Cartridge (purification rack position 1) before a purification run.

Get recall alerts

Free email alert whenever Roche Molecular Systems, Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Roche Molecular Systems, Inc.