SCC Soft Computer recalls SoftPath Laboratory Information System. Versions 4.3.0.8, 4.3.0.9, 4.3.0.10, 4.3.0.11, 4.3.0.12, 4.3.0.14, 4.3.0.15, an…
- Recall date
- November 26, 2014
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1065-2015
- FDA classification
- Class II
- Brand / firm
- SCC Soft Computer
- Sold / distributed
- Distribution US Nationwide and CANADA.
Why it was recalled
Modifications to diagnostic text may be: 1) Saved to the database but not appear on the report sent to the physician; or 2) Documented on the report, but not saved to the database.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
SoftPath Laboratory Information System. Versions 4.3.0.8, 4.3.0.9, 4.3.0.10, 4.3.0.11, 4.3.0.12, 4.3.0.14, 4.3.0.15, and 4.4.0.0