Medical device recalls Moderate risk

Schaerer Medical USA Inc recalls Replacement floor locking cylinders for the Schaerer model 7300 surgical tables. Replacement cylinders are part number…

Recall date
September 27, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0372-2017
FDA classification
Class II
Brand / firm
Schaerer Medical USA Inc
Sold / distributed
Worldwide Distribution - US Nationwide in the states of: CA, CT, FL, IA, IN, MI, MN, MO, OH, PA, and VA Replcement cylinders wre distributed

Why it was recalled

The floor locking cylinders are contained in the base of the Schaerer 7300 model surgical tables. The cylinders were manufactured with the incorrect "O" ring which may cause the cylinder to leak.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Replacement floor locking cylinders for the Schaerer model 7300 surgical tables. Replacement cylinders are part number 85118. Product Usage: The cylinders are used to lock the surgical table in place.

Get recall alerts

Free email alert whenever Schaerer Medical USA Inc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Schaerer Medical USA Inc