Medical device recalls Moderate risk

Scientia Vascular, Inc. recalls Socrates 38 Aspiration Catheter 127 cm length REF SC038-127-001 The Socrates Aspiration System with a compatible suctio…

Recall date
January 2, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1234-2025
FDA classification
Class II
Brand / firm
Scientia Vascular, Inc.
Sold / distributed
U..S. Nationwide distribution in the states of AL, AZ, FL, GA, HI, IL, IN, KS, MN, NC, NJ, NY, PA, SD, TX, VA, WI and WV.

Why it was recalled

Due to manufacturing non-conformance that resulted in channels in the seal on the packaging pouch. As a result, this could potential create a breach in the sterile barrier.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Socrates 38 Aspiration Catheter 127 cm length REF SC038-127-001 The Socrates Aspiration System with a compatible suction pump is intended for use in the revascularization of patients with acute ischemic stroke

Get recall alerts

Free email alert whenever Scientia Vascular, Inc. has a new recall — straight from official government data. Unsubscribe anytime.