Medical device recalls Moderate risk

Seca recalls Seca 402/403 Baby Scale Cart, Ref # 4020000009/ REF # 4030000009

Recall date
June 19, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1429-2020
FDA classification
Class II
Brand / firm
Seca
Sold / distributed
US: CA, MI, WA, WV, VA, NM, LA, TN, IL, PA, OH, FL, WA, OR, NY, CO, IA, MN, TX, MA, NC, LA, KY, NM AL,DC, NJ, AK, AZ, NV, SC, HI, MD, WI, RI, DE, MO OUS: Canada

Why it was recalled

It has been determined that the wheels on the baby scale cart can unexpectedly loosen and in the worst case, detach from the cart completely. If this occurs the cart can become unstable and tip over, potentially leading to the scale falling from the cart. If a baby is left unattended in a cart, the baby could fall off the scale and be subjected to several serious injuries.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Seca 402/403 Baby Scale Cart, Ref # 4020000009/ REF # 4030000009