Securitas Healthcare LLC recalls Arial 900 MHz Call Station, Model Number CSK200-900, with Firmware version: 1.6.3
- Recall date
- December 8, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1031-2026
- FDA classification
- Class II
- Brand / firm
- Securitas Healthcare LLC
- Sold / distributed
- Worldwide distribution -US Nationwide and the country of Canada.
Why it was recalled
Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Arial 900 MHz Call Station, Model Number CSK200-900, with Firmware version: 1.6.3
Get recall alerts
Free email alert whenever Securitas Healthcare LLC has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Securitas Healthcare LLC