Medical device recalls Moderate risk

Securitas Healthcare LLC recalls Arial 900 MHz Call Station, Model Number CSK200-900, with Firmware version: 1.6.3

Recall date
December 8, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1031-2026
FDA classification
Class II
Brand / firm
Securitas Healthcare LLC
Sold / distributed
Worldwide distribution -US Nationwide and the country of Canada.

Why it was recalled

Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Arial 900 MHz Call Station, Model Number CSK200-900, with Firmware version: 1.6.3